Calibration

Calibration

What is Calibration?

Maybe you are not familiar with the meaning of calibration of medical devices, which is an activity to determine the conventional truth of the value of the designation of measuring tools and materials.

This is done by comparing traceable measuring standards to national and international standards. This activity is important for every medical device, especially those that are routinely used (used every day) in health facilities.
Government regulations are very clear in setting the criteria for medical equipment that must be calibrated. The following are the criteria for medical equipment in health care facilities that must be calibrated at least once a year:
  1. Do not have certificates and marks of passing the testing or calibration of medical devices.
  2. The certificate has expired and has passed the test.
  3. The designation or output or performance or safety of the medical device is no longer appropriate, even though the certificate and test pass are still valid.
  4. Medical devices have been damaged and repaired, even though the certificate and test pass are still valid.
  5. Has been transferred from one agency to another, even though the certificate and test pass are still valid.
  6. If the certificate and the test pass are lost or damaged so that information from medical equipment cannot be obtained.

The criteria for medical equipment that have passed calibration and are suitable for use are as follows:

  1. The deviation value of the measurement results when compared with the value enshrined in the medical equipment does not exceed the allowable deviation value.
  2. The work safety measurement value is included in the normal/allowed threshold value.

By performing periodic calibrations, the performance and accuracy of medical devices can be maintained properly. Thus, the quality of health services to patients can also be maintained.

Why Medical Devices Need Calibration?

Calibration of medical devices is an activity that must be carried out, especially when medical equipment has entered the 6 criteria mentioned above. The purpose of this calibration activity is to:

  1. Ensure Conformity of Characteristics SpecificationsThe first objective is to ensure the characteristics of the specifications of an instrument/tool or measuring material. If there is a discrepancy, then the permit or certification letter cannot be issued.
  2. Determine the Deviation ValueCalibration of medical devices also aims to determine the deviation or deviation of the conventional truth from the designation value of the measuring material/instrument, as well as the deviation of the nominal dimensions that should be for a measuring material.
  3. Maintain the Accuracy of the Values ProducedThe next goal is to maintain the accuracy of the value produced from medical devices so that the results do not deviate far from the specified threshold (according to the standard).
  4. Guarantee Standardized Measurement Results: In addition, calibration will also ensure that the measurement results of a medical device are in accordance with national and international standards.
A medical device used for services at a health facility needs to be calibrated from time to time, even though it still has a certificate of eligibility. Especially if the medical equipment is widely used, the feasibility test must always be carried out.
If you want to test various medical devices, you can entrust them to Stark Medical. Stark Medical provides medical equipment calibration services and has obtained a medical equipment test and calibration operation permit from AS3551:2012.
So, you don’t need to hesitate to carry out the calibration process of medical devices and leave the work to professionals to complete it.

    • Related Articles

    • Ergometer Calibration Service

      Overview Ergometers that are used as medical devices must be annually checked with a metrological control (MTK). Ergotest 550 has the highest measurement accuracy required for this and can be calibrated by Stark Medical. Dynamic Ergometer Calibration ...
    • SPIROLAB

      USING THE SPIROLAB TURNING THE SPIROLAB ON AND OFF To turn the spirolab on, press then release the button on the front of the unit. When you turn on the device, the first screenshot shows the internal software loading progress. Only the first time ...
    • Requirements for Medical Practices around Medical Device Yearly Maintenance and Accreditation

      Why are some clinics under accreditation guidelines while others are not? Accreditation is only required for hospitals, day procedure services and public dental services in Australia. These are licenses private hospitals and day procedure centres. ...
    • Stress ECG

      Stress ECG and Cubestress software Cardioline is an Italian based manufacturer solely focused on high tech Diagnostic Cardiology equipment including Resting ECG, Holter, Stress and ABPM. The new wireless Cubestress ECG was released into the ...
    • TouchECG

      Cardioline TouchECG HD+ setup guide